Dual Pass Transmissometer Dust Monitor BI 7000

Dual Pass Transmissometer Dust Monitor BI 7000 is an optical instrument designed to measure the concentration of dust or particulate matter suspended in exhaust gas passing through a duct or stack of industrial process.

Dust Monitor uses the double pass transmission measurement technology. Light beam emitted from the transmitter passes across the stack to a reflector, which then return the light to the transceiver where the intensity of the received light is measured.

When calibrated against standard reference measurements, the amplitude of this signal relates directly to the particulate concentration in the stack, and can be used to present a reading in mg/m³. The advantage of is that it is significantly less susceptible to drift with time, temperature or dirtying optics, than traditional dust monitors.

The BI-7000 has no moving parts and with SS body is of rugged design and has an excellent reliability record. With minimum maintenance and excellent performance BI-7000 dust monitor is best suited for dust concentration measurement.

Transmitter are supplied with an air purge facility which when connected to a high volume source of instrument quality air will resist particle deposition on the lenses for accurate measurement.

Category:

Features 

  • Direct measurement in exhaust flue gas flow – In situ
  • Dual Pass for accurate reading
  • Unit programmable in mg/m3, mg/Nm3, % Opacity
  • RS 485, MODBUS protocol
  • Rugged design with no moving parts so low maintenance
  • Green LED source for long lifetime stability
  • IP65 rating Enclosure
  • The best price performance on the market – designed with user requirements foremost in mind

Specifications

Path Length 1 – 20 m flange to flange separation
Measuring Range 0 -1000 mg/m³ user selectable
Units of measurement mg/m³, mg/Nm³, % Opacity
Accuracy ±1% relative to the maximum range
Resolution 1mg/m³
Drift with Temperature ±1% over any 20°C in the operating range
Operating Wavelength 510 – 540 nm Green LED
Stack Temperature 300° C
Power Supply 110 – 230 VAC
Air Supply 5 – 10 m³/hr Flow & 500 mbar Pressure. Must exceed maximum stack pressure.
Air Inlet Connection 1” BSP threaded
Output 4 – 20 mA Analog output
Digital Interface RS 485, MODBUS protocol
Mounting Wall
Electronic Unit Enclosure Rating IP 65
Tx & Rx Junction Box Rating IP 65
Cable 2 core screened cable of 0.5mm²,
8 meters between Tx & Rx
6 core screened cable of 0.5mm²,
25 meters between Rx & Electronic Unit
Ambient Temperature -20 to 50° C
Operating Humidity 5 – 95%

Applications

  • Aluminium
  • Cement
  • Distillery
  • Fertilizers
  • Iron & steel
  • Oil refinery
  • Petrochemical
  • Power Plants
  • Thermal Power Plants
  • Sugar
  • Tannery
  • Zinc
  • Copper
  • Biomedical waste Incinerator
  • Common Hazardous Waste Incinerator
  • Boiler

Download Materials

Technical Brochure

Application Questionnaire

Product Inquiry

    Contact

    Newsletter

    Stay up to date with our latest analytical solutions, services, and innovative technologies. Don’t miss any updates as we continuously bring new and exciting developments tailored to enhance your operations.

      About Company

      Kite Analytics has been providing cutting-edge analytical solutions since 2022. Specializing in data-driven services for various industries, our focus is on delivering quality and precision to help businesses achieve operational excellence.

      © 2024 Voxitech — Built with Precision & Innovation